A blood pressure is only as good as the way it was taken. Size the cuff from the arm, not the patient, predict which way each common error pushes the number, classify readings by the current adult categories, and read an orthostatic set the way you would at the bedside. Every case is generated, so you can keep practicing.
Size the cuff from the arm
The bladder width should be about 40 percent of the mid upper arm circumference and its length about 80 percent (many guidelines accept 75 to 100 percent). Measure the arm; do not guess from the patient’s size.
Which cuff on the shelf fits?
Which way does this error push the reading?
Each card describes one thing that went wrong. Decide what it does to the number you are about to chart.
Classify this adult reading
Use the 2017 ACC and AHA categories (kept in the 2025 update). If systolic and diastolic fall in different categories, the higher category wins. Then work out the pulse pressure and the MAP.
Read the orthostatic set
Supine for at least 5 minutes, then stand (or sit and dangle if standing is unsafe) and remeasure at 1 minute and 3 minutes. Orthostatic hypotension is a drop of at least 20 systolic or 10 diastolic within 3 minutes.
Fit check
Work out the ideal width and length, choose a cuff, then press Check my sizing.
Bare arm, supported, cuff at heart level (midpoint of the sternum). Avoid an arm with an IV infusion, a dialysis fistula or graft, injury, a cast or paralysis, and follow policy after mastectomy with node removal.
Why it happens
Pick what the error does to the reading.
Pattern to remember. Almost every preparation error reads falsely high: small cuff, loose cuff, arm below the heart, unsupported arm, crossed legs, talking, a full bladder. Only a few read low: a cuff that is too large, an arm above the heart, and the auscultatory gap. Deflating too fast does both: systolic low, diastolic high.
The auscultatory gap. In hypertension the sounds can vanish between phase I and phase II and return. If you start listening inside that silence you chart a falsely low systolic. Palpate first: inflate until the radial or brachial pulse disappears, deflate, wait about 30 seconds, then inflate 20 to 30 mm Hg above that number and deflate at 2 to 3 mm Hg per second.
Category, pulse pressure and MAP
Choose a category and fill in both numbers, then press Check.
Formulas. Pulse pressure = systolic minus diastolic (normally about 40). MAP = (systolic + 2 × diastolic) ÷ 3. A MAP of at least 65 is the usual minimum target for organ perfusion in sepsis and shock.
Interpretation
Look at the drop and the heart rate response, then choose.
Stay at the bedside. Use a gait belt as needed and stop if the patient becomes dizzy or faint, returning them to bed. Orthostatic hypotension is a major cause of falls.